Sterilization time:
1
minute increments within a range of 1-250 minute
Liquefaction temperature: In increments of 1°C within
a
range of
60-
100°C
Liquefaction
time: 1 minute increments within a range of
1-60
minutes
Warming temperature:
1°C
increments with a range of 45-60°C
EXHT
%:
1%
increments within a range of 0-100%
If you hold the switch down, the displayed value increases or decreases
in 10 unit increments. When the displayed value exceeds the upper
limit (lower limit), it returns to the lower limit (upper limit).
@
Press the
SETLENT
switch.
The changed value is stored, the display stops flickering
and
lights
up.
\
I
This completes the setting.
I-
Canceling Changing the
Set
Values
v
If halfway you desire to cancel the setting change, press the
MODE
switch. The changed value is not stored, and the equipment returns to a
standby state.
-
NOTE
For sterilization of liquid, set a sterilization time longer than desired, taking delay time into
account and referring to the table below.
Example)
In
case there is 3 liters of water in a flask, it takes nearly 30 minutes (delay time) for
temperature
of
water in the container to reach a set sterilization temperature after
temperature in the working chamber reaches the set value. You should set a sterilization
time 30 minutes longer than desired to cope with this delay of time. Therefore, the set
sterilization is
50
minutes:
Set sterilization time
(50
minutes)
=
Delay
time (30 minutes)
+
desired sterilization time
(20
minutes)
Set
sterilization
HV-50
Reference Values
of
Delay
Time
.
,
Dehy,time:
~~~=
,
Sterilization
time
(per Flask)
Quantity
of
Liquid
3
liters
2
liters
1
liter
30
minutes
25
minutes
20
minutes
Temperature
in
working
chamber
HV-50
Delay Time Reference Data