Chapter 2. Installation - A CAUTION If the equipment is installed in a place which is 800 m or higher than sea level (i.e. under low pressure in mount

Strany 3

2, Installation Procedure @ Put the body on the caster stoppers to prevent the body from ac Anchor the body as described in the following. (1) Set the

Strany 4

@ Pour water into the exhaust bottle. Add water to the exhaust bottle as described below. Drain po Pol Reference line level (1) UnIoad the exhaust bot

Strany 5

@ Referring to "chapter 3. Operation Method," open the lid and take out the accessories. @ Place the bottom place in the working chamber. @

Strany 6 - HV-25/50185/11 O?"

Chapter 3. Operation Method Basic Operation Method Turning "ON' Power Switch 1 See '1. Turning 'ON' POWER Switch" on pag

Strany 7

1. Turning "ON" POWER Switch O Turn ON the POWER switch at the front of the body. When the openiclose lever is set to "LOCK" (lef

Strany 8 - HV-25/50/85111 O?

3. Loading Substance I Be careful not to pinch hands when closing the lid. Check to see that the area of the lid that is to come in contact with the l

Strany 9 - INCREASEDECREASE

Do not pile specimens on top of one another. When the chamber is overly packed, steam fails to penetrate to all points, resulting in incomplete steril

Strany 10 - Installation

Mode 1 Mode 2 Mode 3 Mode 4 Initial Set Value XI-J- Step Display WEATG -+ STER. -+ EXHT. (Varied) -+WARM Step Display BEATG -+ STEW. -+ EXHT. (Varied)

Strany 11 -

5. Changing Set Values (Registering Values by Customer) To change set values (sterilization temperature, sterilization time, warming temperature, open

Strany 12 - Drain po

Introduction We would like to express our gratitude for your purchase of our autoclave. This brochure covers the operation method and a simple mainten

Strany 13 - (for receiving steam)

@ Sterilization time: 1 minute increments within a range of 1-250 minute Liquefaction temperature: In increments of 1°C within a range of 60- 100°C L

Strany 14 - Operation

I If steam is abruptly exhausted after sterilization of liquid, the liquid may gush out. To prevenl this, set the opening of the exhaust valve (Exha

Strany 15 - <r>

6. Checking and Correcting the Clock Time @ With the equipment in a standby state, press the FUNC. switch twice. The CILOCK display lights up, and the

Strany 16 - IMPORTANT

8 Press the SETIENT switch. The corrected current time is stored, and the equipment returns to a standby state. Canceling Correcting Clock If halfway

Strany 17 - (Process)

7. Checking and Setting of Turn-on Timer a IMPORTANT 0 Invalid setting of the turn-on timer and an incorrect clock may lead to equipment i malf

Strany 18 - (Fixed)

Example of Setting: When you desire to start operation at 11:30 p.m. today, your setting should be as follows: Day(s)-later setting = 0, Hour settin

Strany 19 - Customer)

8. Starting Operation O Check again to see that the water level in the exhaust bottle is between the HIGH and LOW levels. ... If above the HIGH level

Strany 20 - Dehy,time:

- Exhaust % During the Valve Opening Variable Exhaust Cycle O Press the SETIENT switch. The display of opening of the exhaust % display starts flicker

Strany 21

9. Unloading A WARNING Check to see that the gauge for pressure in the working chamber reads 0 kgf/cm2. - A CAUTION Do not put your face or hands c

Strany 22 - Checking

11. To Interrupt Operation A NOTE If operation is interrupted during the sterilization of liquid, the liquid may overflow into the worlung chamber. I

Strany 23 - SETIENT

Before Using In this manual the following headwords are applied to items to which great attention should be given: WARNING: Precaution indicating an

Strany 24

Any air remaining in the working chamber makes the temperature distribution in the chamber uneven. This hinders temperature rise (sterilization). To c

Strany 25 - Tl[MIEK

When the pressure in the working chamber falls below 0.1 kgf/cm2, the exhaust valve is fully opened. When the temperature in the working chamber

Strany 26

Completion Cycle --- Common to all the Modes @When all cycles of each mode are complete, the electronic alarm indicates completion of all cycles

Strany 27 - 1-100%

Chapter 4. Maintenance and Service Be sure to start maintenance or service work after the main body has been sufficiently cooled. r I cAuT1oN 1. Drain

Strany 28 - WARNING

-- - - - - - --- 2. Draining Working Chamber When you plan to put the equipment out of service for a long time, be sure to drain the working chamber t

Strany 29 - Unloading"

3. Body Service , a IMPORTANT Do not use benzine or thinner to clean the body. Also make sure that the volatile substances such as insecticides do not

Strany 30 - Temperature

(2) Push in the gasket unit it hits against the metal surface. When the gasket is @- (2) d- Flat porton completely pushed in, the flat portion

Strany 31

Chapter 5. Specification Outside dimensions HV-50 Model Inside dimensions Rated power supply Temperature and humidity conditions HV-25 1 480W x 950H x

Strany 32 - Completion

I Model I Outside dimensions Inside dimensions Rated power supply 420 rnrn dia. x 615 mm depth I (Effective: 85 liters) 420 mrn dia. x 795 rnm depth (

Strany 33 - Maintenance and Service

Piping Diagram Pressure Pressure gauge sensor Working chamber I Drain valve - Safety valve Exhaust valve 41 I Exhaust bottle fl

Strany 34 - CAUTION

(5) Harnmable gas (hydrogen, acetylene, ethylene, methane, ethane, propane, butane, and other substances that are gas at a temperature of 15°C at 1 at

Strany 35

Wiring Diagram UUL DP CP SW ELB m F B M Disphy printed circuit board Control printed circuit board Operation switch Power switch Transformer Fuse Back

Strany 36

Chapter 6. In the Event of Suspected Failure 1. Error Detection (Alarming) Operation Should any malfunction occur in your autoclave, the incorporate

Strany 37 - Specification

2. Early Trouble .Shooting Symptom Display remains off after turning on power. No air exhausted from working chamber. Pressure gauge reading remains l

Strany 38

Temperature-Pressure Conversion Table for Saturated Steam in Autoclave The following data are approximate values for practical use in all autoclaves:

Strany 39 - Diagram

SPARE PARTS FOR HV- AND HVE-SERIES AUTOCLAVES Catalog No. ATA-25LG ATA-50LG ATA-85LG ATA-25H ATA-50H ATA-85H ATA-11 OH ATA-HVEGO - ATA-HVEROM ATA-HVES

Strany 41 - Failure

Looking for more information?Visituson thewebat http://www.artisan-scientific.com for more information:•PriceQuotations •Drivers·TechnicalSpecificatio

Strany 42 - .Shooting

Do not unload the exhaust bottle or drain the working chamber when the chamber is under pressure. Boiling water or steam may gush out, and you may ge

Strany 43 - (OC) (OF) kglcm lblinch

How to Read This Manual a This operation manual consists of the following sections covering the information required for proper operation of the autoc

Strany 44 - SPARE PARTS

Contents Introduction ... .. ... Table of Con

Strany 45

Cha~ter 1. What Is Autoclave HV-25/50/85111 O? 1. Product Uses This product is designed for research use only (Modes 1 - 3). It is not labeled or mark

Strany 46 - • Equipment Demos

Display and Operation Switches @ Digital Display (Temperature, Time,of Day, Error) The digital display indicates set temperature, current time and pre